Quality Control

Development of the national standard material of progesterone in frozen human serum*

Expand
  • 1. National Institutes for Food and Drug Control, Beijing 100050, China;
    2. Reference System Center of Maccura Biotechnology Co., Ltd., Chengdu 611731, China

Received date: 2023-04-29

  Online published: 2024-06-21

Abstract

Objective: To develop the national standard materials of progesterone in frozen human serum and to evaluate the accuracy and standardization of serum progesterone detection kits. Methods: Low and high concentrations of serum were from normal people and pregnant women individuals respectively, with a clear appearance, no obvious hemolysis, jaundice and lipid blood. After all tests for four infectious diseases were negative, thrombin, anhydrous calcium chloride and anhydrous sodium carbonate were added according to the proportion of 1 kU·L-1, 2.219 6 g·L-1 and 2.65 g·L-1 respectively, and centrifuged (4 000 r·min-1, 4 ℃, 25 min) after fully mixing. After twice of multiple filtration and sterilization, serum pools were packed in ampoules to prepare candidates Ⅰ and Ⅱand stored at -70 ℃. Single factor variance method and linear regression method were used to evaluate the homogeneity and stability of the candidate. Five labs working together used reference method (isotope dilution liquid chromatography tandem mass spectrometry) to assign value for candidate. The uncertainty was calculated. At the same time, the commutability was evaluated. Results: Through statistical analysis, the F values of homogeneity for candidate Ⅰ and Ⅱ were 0.570 9 and 1.200 9, respectively, which were all less than F0.05. Candidate Ⅰ and Ⅱ could be stable at least 8 h at 20-25 ℃. At 2-8 ℃, candidate Ⅰ and Ⅱ could be stable at least 33 d and 7 d respectively. Candidate Ⅰ and Ⅱ cound be stable at least 33 d at -20 ℃. Candidate Ⅰ and Ⅱ remained stable after repeated freeze-thaw cycles for 6 times. The assigned values were: candidate Ⅰ: (4.44±0.24) ng·mL-1 (k=2), Ⅱ: (16.79±0.82) ng·mL-1 (k=2). The determination results of candidate Ⅰ and Ⅱ were all within the 95% confidence interval of the fitting regression line established by the determination values of fresh serum samples. Conclusion: Progesterone candidate had good homogeneity, stability, interoperability. It had accurate and reliable determination value, which could be used as national standard. The establishment of this national standard was of great significance for promoting the consistency and standardization of progesterone testing results.

Cite this article

YU Ting, TANG Li-ping, HUANG Jie, SUN Ke-qi . Development of the national standard material of progesterone in frozen human serum*[J]. Chinese Journal of Pharmaceutical Analysis, 2024 , 44(3) : 547 -552 . DOI: 10.16155/j.0254-1793.2024.03.21

References

[1] MOREL Y,ROUCHER F, PLOTTON I, et al. Evolution of steroids during pregnancy: maternal, placental and fetal synthesis[J]. Ann Endocrinol (Paris), 2016, 77(2): 82
[2] SINGH B,RESCHKE L,BAKER VL. Frozen-thawed embryo transfer:the potential importance of the corpus luteum in preventing obstetrical complications[J]. Fertil Steril, 2020, 113(2):252
[3] 云永子,刘宇杰,陶阳,等.经阴道超声联合腹部超声诊断异位妊娠临床价值分析[J]. 医学影像学杂志,2022,32(1):172
YUN YZ, LIU YJ, TAO Y, et al. Clinical value of tmnsvaginal ultmsound combined with abdominal ultrasound in the diagnosis of ectopic pregnancy[J]. J Med Imag, 2022,32(1):172
[4] 胡羽晶,任叶丹,黄云柯. 血清单核细胞趋化蛋白-1、β-人绒毛膜促性腺激素、孕酮及雌二醇水平预测高龄先兆流产患者保胎失败的价值探讨[J]. 中国性科学, 2022, 31(7):103
HU YJ, REN YD, HUANG YK. Value of serum MCP-1,β-HCG,P and E2 level in predicting failed fetal protection in elderly patients with threatened abortion[J]. Chin J Human Sex, 2022, 31(7):103
[5] 陈海波,王婷,杨达刚. 血清孕酮、β-HCG及尿碘水平对早期妊娠先兆流产保胎结局的预测价值[J]. 贵州医科大学学报, 2022, 47(6):717
CHEN HB, WANG T, YANG DG. Predictive value of the levels of serum progesterone, β-HCG, and urinary iodine on the fetus outcomes after threatened abortion in early pregnancy [J]. J Guizhou Med Univ, 2022, 47(6):717
[6] JEWSON M,PUROHIT P,Lumsden MA. Progesterone and abnormal uterine bleeding/menstrual disorders[J]. Best Pract Res Clin Obstet Gynaecol, 2020, 6(19):62
[7] 陈超,任程程,刘峰,等. 一种血清的制备方法:中国,201610792330.8 [P]. 2017-02-15
CHEN C, REN CC, LIU F, et al. A method for preparing serum: China, 201610792330.8 [P]. 2017-02-15
[8] 国家质量监督检验检疫总局. JJF 1343-2022标准物质的定值及均匀性、稳定性评估 [S]. 2022
General Administration of Quality Supervision, Inspection and Quarantine of the People's Republic of China. JJF 1343-2022 Characterization, Homogeneity and Stability Assessment of Reference Material [S]. 2022
[9] SUSAN S-C T, BEI X, MICHAEL J WELCH. Development and evaluation of a candidate reference measurement procedure for the determination of progesterone in human serum using isotope-dilution liquid chromatography/ tandem mass spectrometry[J]. Anal Chem, 2006, 78(18): 6628
[10] 中华人民共和国卫生部 WS/T356-2011 基质效应与互通性评估指南[S]. 2011
Ministry of Health of the P.R. China. WS/T356-2011 Guideline for Evaluation of Matrix Effects and Commutability[S]. 2011
[11] 黄晓楠,刘玉林. 全自动化学发光分析仪检测孕酮和人绒毛膜促性腺激素的性能验证[J]. 医学信息, 2022,35(11):134
HUANG XN, LIU YL. Performance verification of automatic chemiluminescence analyzer for detection of progesterone and human chorionic gonadotropin[J]. J Med Inf, 2022,35(11):134
[12] 程跃建. 梅里埃VIDAS免疫荧光分析仪、贝克曼DXI-800化学发光仪检测血清β-HCG、孕酮结果一致性分析[J]. 医学检验与临床, 2018, 29(11): 11
CHENG YJ. The condidtency anslysis of serum β-HCG and progesterone measured by VIDAS imunofluorescence analyzer and Backman DXI-800 chemiluminescence analyzer [J]. Med Lab Sci Clin, 2018, 29(11): 11
[13] 曾帮智,杨海蓉. 血清IAP、孕酮、雌二醇及PAPP-A水平对妊娠早期胚胎停止发育的预测价值[J]. 医学临床研究, 2023,40(8): 1161
ZENG BZ, YANG HR. Predictive value of serum IAP, progesterone, estradiol, and PAPP-A levels in early pregnant em-bryo arrest[J]. J Clin Res, 2023,40(8): 1161
[14] 李洁. FT4、FT3、E2、孕酮水平检测在乳腺癌诊断中的价值[J]. 中国民康医学, 2023,35(14): 117
LI J. Value of FT4, FT3, E2 and P levels in diagnosis of breast cancer [J]. Med J Chin People's Health, 2023,35(14): 117
[15] 欧阳芬,张乔轩,严君,等. 血清孕酮同位素稀释液相色谱串联质谱候选参考方法的性能评价及临床应用[J]. 中华检验医学杂志, 2022,45(5): 456
OUYANG F, ZHANG QX, YANG J, et al. Performance evaluation of serum progesterone measurement by ID-LC/MS/MS candidate reference methods and their clinical application value [J]. Chin J Lab Med, 2022,45(5): 456
Outlines

/