Safety Monitoring

Investigation the container closure integrity of the pressurized metered dose inhalers by helium mass spectrometry

Expand
  • 1. School of Chemical Engineering, University of Shanghai for Science and Technology, Shanghai 200093, China;
    2. Shanghai Shangyao Xinyi Pharmaceutical Co., Ltd, Shanghai 201206, China;
    3. Hangzhou Chance Pharmaceuticals Co., Ltd, Hangzhou 311121, China

Received date: 2023-11-16

  Online published: 2024-06-21

Abstract

Objective: To establish and verify the helium mass spectrometry detection method to test the sealing integrity of the container closure system (CCS) of pressurized metered dose aerosols (pMDIs). Methods: Determined the maximum allowable leakage limit (MALL) of pMDIs of CCS in the whole life cycle. The special test chambers and helium filling devices were designed and manufactured, and helium filling devices to test the negative and positive control samples under high pressure (absolute pressure 672 kPa) and normal pressure (absolute pressure 100 kPa) respectively. According to the requirements of methodology validation, the validation of detection limit, system suitability, precision, specificity and detection range indicators were completed.The accelerated sample sealing integrity testing for 3, 6 months and 3, 12, 24 months in the long-term stability inspection period was achieved. Results: Under both helium-filled pressure conditions, the methodological validation indicators were in line with the acceptable standards, and the results of the stability samples were less than the detection limits. This method can detect 100% that the equivalent pore diameter of pMDIs product CCS was 0.095 μm and above. Conclusion: The helium mass spectrometry detection method can quickly and quantitatively investigate the leakage rate of pMDIs products, prove the seal integrity of CCS with high sensitivity, and meet the industry's requirements for pMDIs product quality control.

Cite this article

LEI Xiao-qing, SHAO Qi, CHEN Lan . Investigation the container closure integrity of the pressurized metered dose inhalers by helium mass spectrometry[J]. Chinese Journal of Pharmaceutical Analysis, 2024 , 44(2) : 298 -306 . DOI: 10.16155/j.0254-1793.2024.02.13

References

[1] MYRDAL PB, SHETH P, STEIN SW, et al. Advances in metered dose inhaler technology: formulation development[J]. AAPS PharmSciTech, 2014, 15(2):434
[2] STEIN SW, SHETH P, HODSON PD, et al. Advances in metered dose inhaler technology: hardware development[J]. AAPS PharmSciTech, 2014, 15(2):326
[3] ZULEGER B, WERNER U, KORT A, et al. Container/closure integrity testing and the identification of a suitable vial/stopper combination for low-temperature storage at -80 ℃[J]. PDA J Pharm Sci Technol, 2012, 66(5):453
[4] 翁琳琳, 唐黎明, 陈桂良, 等. 吸入给药制剂的质量控制及其安全性研究进展[J]. 药物分析杂志, 2013, 33(5):724
WENG LL, TANG LM, CHEN GL, et al. The progress of the research on quality control and safety evaluation of inhalation preparations[J]. Chin J Pharm Anal, 2013, 33(5):724
[5] 中华人民共和国药典 2015 年版. 四部[S]. 2015: 70
ChP 2015. Vol Ⅳ [S]. 2015: 70
[6] USP 40 [S]. 2017:354
[7] 国家药品监督管理局药品审评中心. 化学药品注射剂仿制药质量和疗效一致性评价技术要求[EB/OL]. 2020 [2022-02-09]. https://www.cde.org.cn/main/news/viewInfoCommon/d9c6f118b773f54e8feba3519bf78a11
Center for Drug Evaluationm, NMPA. Technical Requirements for Quality and Efficacy Consistency Evaluation of Generic Drugs for Chemical Injections[EB/OL]. 2020 [2022-02-09]. https://www.cde.org.cn/main/news/viewInfoCommon/d9c6f118b773f54e8feba3519bf78a11
[8] 国家药品监督管理局药品审评中心. 化学药品注射剂包装系统密封性研究技术指南(试行) [EB/OL]. 2020 [2022-02-09]. https://www.cde.org.cn/main/news/viewInfoCommon/8a4f9f16844fbed617f8e8ed59485c1d
Center for Drug Evaluation,NMPA. Guideline to container-closure integrity test for chemical injections (Trial) [EB/OL]. 2020 [2022-02-09]. https://www.cde.org.cn/main/news/viewInfoComm-on/8a4f9f16844fbed617f8e8ed59485c1d
[9] 马骏威, 郭涤亮, 刘涓, 等. 胶塞/玻璃瓶容器密封完整性研究的探讨[J]. 中国新药杂志, 2021, 30(5):5
MA JW, GUO DL, LIU J, et al. Discussion on the container closure integrity of rubber stopper/glass container[J]. Chin J New Drugs, 2021, 30(5):5
[10] Sterile Product Packaging-Integrity Evaluation[S]. 2020:40
[11] 封二飞. 无菌药品包装完整性研究[J]. 中国医药工业杂志, 2019, 50(11):1352
FENG EF. Research of sterile products packaging integrity[J]. Chin J Pharm, 2019, 50(11): 1352
[12] PELÁEZ SS, MAHLER HC, HERDLITSCHKA C, et al. Comparing physical container closure integrity test methods and artificial leak methodologies[J]. PDA J Pharm Sci Technol, 2019, 73(3):220
[13] BROWN H, MAHLER HC, MELLMAN J, et al. Container closure integrity testing—practical aspects and approaches in the pharmaceutical industry[J]. PDA J Pharm Sci Technol, 2017, 71(2):147
[14] USP. General Chapter 1207.2. Package Integrity Leak Test Technologies [S]. 2016
[15] 陆维怡, 姚雪凌, 孙甜甜. 氦质谱(真空模式)考察管制玻璃注射剂瓶包装密封完整性[J]. 药物分析杂志, 2022, 42(4):704
LU WY, YAO XL, SUN TT. Helium mass spectrometry (vacuum mode) used to investigate the container closure integrity of the vials system for injection[J]. Chin J Pharm Anal, 2022, 42(4):704
[16] ASTM F 2391-05(2016). Measuring Package and Seal Integrity Using Helium as the Tracer Gas[S]. 2016
[17] 肖祥正. 泄漏检测方法与应用[M]. 北京: 机械工业出版社, 2010: 5
XIAO XZ. Leak Detection Method and Application[M]. Beijing: China Machine Press, 2010:5
[18] GLAXO WELLCOME SA. 葛兰素史克公司原研硫酸沙丁胺醇说明书[Z]. 2022:2
GLAXO WELLCOME SA. Instructions Forsalbutamol Sulfate Originally Developed by Glaxo Smith Kline[Z]. 2022:2
[19] CHEN SC. Container closure integrity testing of primary containers for parenteral products[J]. Chall Prot Prod Dev, 2018, 68(5):253
[20] 张景辰, 邵奇, 胡玥, 等. 基于风险的定量吸入气雾剂包装容器系统相关研究要点[J]. 中国医药工业杂志, 2019, 50(7):722
ZHANG JC, SHAO Q, HU Y, et al. Risk based key points of research on packaging container system of metered dose inhaler[J]. Chin J Pharm, 2019, 50(7):722